Pulmonary Vein Isolation With Contact-Force-Sensing Large-Tip Focal Pulsed Field Ablation Catheter With 3D Mapping for Paroxysmal Atrial Fibrillation: Omny-IRE 12-month Results
Back to Research Hub
  1. Home
  2. Research Hub
  3. Pulmonary Vein Isolation With Contact-Fo…

Pulmonary Vein Isolation With Contact-Force-Sensing Large-Tip Focal Pulsed Field Ablation Catheter With 3D Mapping for Paroxysmal Atrial Fibrillation: Omny-IRE 12-month Results

Heart Rhythm·October 22Open Access
Cardiac & Cardiovascular SystemsConfirms priorParoxysmal Atrial FibrillationProspective Single-Arm TrialCatheter AblationPulsed Field AblationAdultOmnypulse

Summary

View source

What was studied

The Omny-IRE trial (NCT05971693) evaluated the safety and effectiveness of the OMNYPULSE platform — a large-tip focal, multielectrode, contact-force-sensing pulsed field ablation (PFA) basket catheter with 3D electroanatomic mapping integration — for pulmonary vein isolation (PVI) in paroxysmal atrial fibrillation (AF), with follow-up through 12 months.

Key findings

In the 136-patient main-phase cohort, 12-month freedom from symptomatic/asymptomatic atrial arrhythmia (≥30 s) was 82.5% (90.1% with standard-of-care monitoring only); clinical success (freedom from symptomatic arrhythmia and acute procedural failure) was 89.9%; device/procedure-related serious adverse events occurred in 2.9% of patients; AFEQT scores improved by 25.5 points; cardiovascular hospitalizations fell 40.9%; and Class I/III AAD use fell 49.3%.

Study limitations

The main-phase cohort was limited to 136 of 188 enrolled patients, potentially introducing selection bias. There was no randomized control arm, limiting causal comparisons. Intensive remote rhythm monitoring (weekly 1–5 months, monthly 6–12 months) may have detected more arrhythmia recurrences than typical real-world practice, explaining the gap between comprehensive and standard-of-care monitoring success rates (82.5% vs 90.1%).

Clinical implications

The OMNYPULSE contact-force-sensing PFA catheter achieved high rates of freedom from atrial arrhythmia recurrence at 12 months with a low serious adverse event rate (2.9%), supporting its use as a safe and effective option for PVI in paroxysmal AF. Clinicians should note that intensive rhythm monitoring protocols will capture more recurrences than standard-of-care monitoring, which matters when counseling patients on expected outcomes.

Caveats

  • Intensive rhythm monitoring (weekly transmissions months 1–5, monthly months 6–12, plus symptom-triggered recordings) is more rigorous than typical clinical practice, meaning real-world recurrence rates may appear lower when using standard monitoring alone (as reflected in the 82.5% vs. 90.1% success rate discrepancy)
  • No randomized control arm was included; effectiveness and safety comparisons to alternative ablation technologies cannot be directly drawn from this study
  • The main-phase cohort (n=136) represents a subset of the 188 enrolled patients; reasons for exclusion of the remaining 52 patients are not detailed in the abstract, which may affect generalizability

Related Questions

Explore related topics

What are the 12-month outcomes of pulsed field ablation versus radiofrequency ablation for paroxysmal atrial fibrillation?How does contact-force sensing improve safety and efficacy in catheter ablation of atrial fibrillation?What is the impact of pulsed field ablation on quality of life and antiarrhythmic drug use in atrial fibrillation patients?