The Omny-IRE trial (NCT05971693) evaluated the safety and effectiveness of the OMNYPULSE platform — a large-tip focal, multielectrode, contact-force-sensing pulsed field ablation (PFA) basket catheter with 3D electroanatomic mapping integration — for pulmonary vein isolation (PVI) in paroxysmal atrial fibrillation (AF), with follow-up through 12 months.
In the 136-patient main-phase cohort, 12-month freedom from symptomatic/asymptomatic atrial arrhythmia (≥30 s) was 82.5% (90.1% with standard-of-care monitoring only); clinical success (freedom from symptomatic arrhythmia and acute procedural failure) was 89.9%; device/procedure-related serious adverse events occurred in 2.9% of patients; AFEQT scores improved by 25.5 points; cardiovascular hospitalizations fell 40.9%; and Class I/III AAD use fell 49.3%.
The main-phase cohort was limited to 136 of 188 enrolled patients, potentially introducing selection bias. There was no randomized control arm, limiting causal comparisons. Intensive remote rhythm monitoring (weekly 1–5 months, monthly 6–12 months) may have detected more arrhythmia recurrences than typical real-world practice, explaining the gap between comprehensive and standard-of-care monitoring success rates (82.5% vs 90.1%).
The OMNYPULSE contact-force-sensing PFA catheter achieved high rates of freedom from atrial arrhythmia recurrence at 12 months with a low serious adverse event rate (2.9%), supporting its use as a safe and effective option for PVI in paroxysmal AF. Clinicians should note that intensive rhythm monitoring protocols will capture more recurrences than standard-of-care monitoring, which matters when counseling patients on expected outcomes.