In-Vitro Comparison of Cyclic Aspiration and Superbore Aspiration Catheters to Conventional 6Fr Aspiration for LVO Stroke

Clinical Neuroradiology·September 16Open Access
Clinical NeurologyLimited evidenceAcute Ischemic StrokeLarge Vessel Occlusion StrokeIn Vitro / Bench StudyAspiration ThrombectomyCyclic AspirationMechanical ThrombectomySuperbore Aspiration CatheterMillipede88React 71Cyclic AspirationSuperbore Aspiration Catheter

Summary

View source

What was studied

This in-vitro study compared three aspiration thrombectomy approaches for LVO stroke — conventional aspiration (CA) using a 6Fr React 71 catheter, Superbore Aspiration Catheter (SAC: Millipede88, 0.088″ lumen), and Cyclic Aspiration (CycA: React 71 with proprietary cyclic pressure waveform) — across soft, mixed, and stiff ovine blood clots lodged in ICA and MCA-M1 segments, using First Pass Effect (FPE), Complete Clot Ingestion (CCI), Final Complete Revascularization (mTICI 2c/3 within 3 passes), and distal emboli rates as endpoints.

Key findings

SAC outperformed both CA and CycA: MCA FPE 88% vs. 71% (CA) vs. 62% (CycA); ICA Final Complete Revascularization 91% vs. 79% for both CA and CycA; and SAC had the fewest distal emboli overall. CycA improved CCI over CA in ICA occlusions (26% vs. 9%) but generated the most distal emboli, particularly in ICA, and did not improve Final Complete Revascularization.

Study limitations

Results come entirely from an in-vitro model — safety outcomes (hemorrhage, perforation) cannot be assessed, and the model lacks M3/M4 vessels, limiting emboli assessment and final tissue revascularization scoring. Only one device per class was tested (Millipede88, React 71, one CycA system), so findings should not be generalized to all superbore or cyclic aspiration devices.

Clinical implications

In this bench model, superbore aspiration (Millipede88) achieved the highest rates of first-pass revascularization and the lowest distal emboli for both ICA and MCA occlusions, supporting its use as a frontline aspiration strategy where anatomy allows. Cyclic aspiration improved clot ingestion in ICA occlusions but increased distal embolization, suggesting adjuncts like balloon guide catheters should be evaluated before broad clinical adoption.

Caveats

  • Major conflicts of interest are present: most authors hold stock options in Perfuze, the manufacturer of the Millipede88 SAC tested, and the lead author has extensive industry relationships including as Principal Investigator for the MARRS-Perfuze study
  • Only one device per technique class was tested; results for Millipede88, React 71, and the proprietary CycA system may not generalize to other superbore or cyclic aspiration catheters
  • Statistical comparisons are exploratory for clot-subtype analyses and the study was not powered for subgroup differences; non-significant subgroup findings should be interpreted with caution
Show all 5
  • The model lacks M3/M4 vessels, causing distal emboli to exit the system rather than lodge distally as in clinical practice, likely underestimating true embolic burden and overestimating final revascularization rates
  • This is a bench (in-vitro) study using ovine blood clots in a silicone model — direct translation to clinical outcomes in human patients is not established

Related Questions

Explore related topics

What are the clinical outcomes of superbore aspiration catheters (≥0.088″) versus conventional 6Fr aspiration in LVO stroke thrombectomy?How does cyclic aspiration thrombectomy compare to standard aspiration in terms of first pass effect and distal embolization in acute ischemic stroke?What is the evidence for mTICI 2c versus 2b as the target reperfusion goal in mechanical thrombectomy for large vessel occlusion?

Publication Details

Year
2026
Journal
Clinical Neuroradiology
Sample Size
n=204