Whether real-time feedback during bag-valve-mask ventilation improves adherence to guideline-recommended tidal volume (400–600 mL) in adults with out-of-hospital cardiac arrest managed by basic life support teams under 30:2 CPR, comparing two sequential cohorts (96 without feedback vs. 67 with feedback).
Real-time feedback improved insufflated volume (Vi) adherence to the 400–600 mL range (68% vs. 42%) and reduced Vi dispersion by 35%, but expiratory tidal volume (Vt) adherence was nearly unchanged (14% vs. 15%), and median Vt was not significantly different (235 vs. 243 mL); leak volume was 12% lower with feedback (mean ratio 0.88, 95% CI 0.77–0.99).
Before-and-after design cannot establish causation; Vt remained well below guideline targets in both periods, suggesting large mask leaks that feedback did not correct; findings from a single Paris Fire Brigade BLS system may not generalize to other EMS settings.
Real-time ventilation feedback helps rescuers squeeze the bag more consistently within target range, but clinicians should know that most of that volume never reaches the patient due to mask leak — improving Vi alone is unlikely to be enough without also addressing mask seal.