Platelet function was evaluated after whole blood (n=5 donors) was circulated for 360 minutes through two total artificial hearts — the Realheart® TAH (RH) and the FDA-approved SynCardia TAH (Sync) — using a standardized ASTM F1830 in vitro protocol (5 L/min, 100 mmHg, 37°C), with matched static controls.
Neither device caused increased platelet activation (CD62P), thrombogenicity (CD61+ microparticles), or mitochondrial damage (ATP, JC-1); SynCardia specifically increased mean platelet volume and reduced ATP levels, suggesting altered particle composition, while platelet function by agonist stimulation and thromboelastometry remained preserved across both devices.
Small sample size (n=5 human blood donors) limits statistical power; the static control exhibited non-physiological changes (higher pH, increased lactate, reduced surface receptor expression), complicating direct comparisons; in vitro testing may not fully replicate the hemodynamic and biological complexity of in vivo implantation.
Clinicians evaluating TAH hemocompatibility should recognize that hemolysis metrics alone do not capture platelet biology; assessing platelet function with metabolic, phenotypic, and functional assays provides a more complete picture, and both devices appear to preserve platelet integrity under controlled conditions.