Functionality of platelets after in vitro testing of pulsatile total artificial hearts

Platelets·October 15Open Access
HematologyLimited evidenceHeart FailureThrombogenicityIn Vitro Experimental StudyTotal Artificial HeartAdultRealheartSynCardia

Summary

View source

What was studied

Platelet function was evaluated after whole blood (n=5 donors) was circulated for 360 minutes through two total artificial hearts — the Realheart® TAH (RH) and the FDA-approved SynCardia TAH (Sync) — using a standardized ASTM F1830 in vitro protocol (5 L/min, 100 mmHg, 37°C), with matched static controls.

Key findings

Neither device caused increased platelet activation (CD62P), thrombogenicity (CD61+ microparticles), or mitochondrial damage (ATP, JC-1); SynCardia specifically increased mean platelet volume and reduced ATP levels, suggesting altered particle composition, while platelet function by agonist stimulation and thromboelastometry remained preserved across both devices.

Study limitations

Small sample size (n=5 human blood donors) limits statistical power; the static control exhibited non-physiological changes (higher pH, increased lactate, reduced surface receptor expression), complicating direct comparisons; in vitro testing may not fully replicate the hemodynamic and biological complexity of in vivo implantation.

Clinical implications

Clinicians evaluating TAH hemocompatibility should recognize that hemolysis metrics alone do not capture platelet biology; assessing platelet function with metabolic, phenotypic, and functional assays provides a more complete picture, and both devices appear to preserve platelet integrity under controlled conditions.

Caveats

  • Extremely small sample size (n=5 blood donors) limits generalizability and statistical conclusions
  • In vitro circulation does not replicate in vivo physiology; static controls showed non-physiological metabolic changes, making direct group comparisons difficult to interpret
  • Results reflect acute 360-minute exposure only; long-term platelet effects of device implantation are not addressed